![Hioki IM3570 Instruction Manual Download Page 90](http://html1.mh-extra.com/html/hioki/im3570/im3570_instruction-manual_2137305090.webp)
4.3 Setting DC Resistance Measurement
82
In cases such as the following, set the time to wait before starting DC resistance measurement.
• When measurement is changed from AC signal measurement to DC resistance measurement.
• When the voltage level is changed.
• When DC adjustment is enabled.
This delay time is for delaying measurement until the DC level and 0 VDC has stabilized.
When measuring DC resistance, observe the measurement waveform and then insert the delay time required
until the transient phenomenon disappears because the time required until the DC level stabilizes differs
depending on the test sample.
4.3.7 Setting the Delay Time
When number of averaging times is 1
Setting changed
•
Changed from AC measure-
ment to DC measurement
•
Signal level changed
•
Range changed
DC delay
(Initial value: 0.0003
seconds)
Measure DC
resistance
DC delay
(Initial value: 0.0003
seconds)
Measure offset
When DC adjustment is enabled (ON)
End measurement
When the number of averaging times is 2 or more
(The number of times is 2 in this example)
Setting changed
•
Changed from AC mea-
surement to DC mea-
surement
•
Signal level changed
•
Range changed
DC delay
(Initial value:
0.0003 seconds)
Measure DC
resistance
DC delay
(Initial value:
0.0003 seconds)
Measure offset
When DC adjustment is enabled (ON)
DC delay
(Initial value:
0.0003 seconds)
Measure DC
resistance
Measure DC
resistance
DC delay
(Initial value:
0.0003 seconds)
Measure
offset
Measure
offset
When DC adjustment is enabled (ON)
Obtain
average
value
Measurement for set number
of averaging times
End mea-
surement
Measurement for set number
of averaging times
Procedure
LCR Initial Screen
1
Rdc Basic Settings
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......