4.2 Setting Basic Settings of Measurement Conditions
46
• When Low Z high accuracy mode (p. 54)
Measurement frequency
setting range
Open circuit voltage
setting range
Open circuit voltage
accuracy
4 Hz to 100 kHz
0.005 V to 1 V
±10%±10 mV
4 Hz
5 MHz
100 kHz
0.005 V
1.000 V
Testing some types of sample is not possible using constant voltage. In this case, the following
symbol appears on the display:
In such a case, constant voltage testing is not performed. Change the constant voltage level
to a value not more than the value being shown as Vmoni.
When a 1
F capacitance is measured at 10 kHz, the CV operation range can be obtained
as follows. Sample impedance
Zm
becomes as follows:
Sample impedance
Zm
becomes as follows:
The impedance
Zm
' observed from the generator is as follows:
Accordingly, the voltage
Vm
across both leads of the sample is as follows:
Because the generator output voltage range is 5[mV] to 5[V] for 10 kHz, the CV operation
range per the above expression is
Vm
= 0.8[mV] to 0.78[V].
In low Z high accuracy mode, the output resistance
Ro
becomes 10[
].
• The open circuit voltage that can be set differs depending on the frequency.
•
When the test frequency is higher than 1 MHz at an open-circuit voltage of more than 1 V, the
voltage is automatically set to 1 V.
• When the measurement frequency is set higher than 1 MHz at an open circuit voltage of less than 0.010 V,
the voltage is automatically set to 0.010 V.
Xm
1
–
2
fC
-----------------
=
Zm
=
Rm
+
jXm
=
0
[
] -
j15.9
[
]
Ro
: Output resistance (100 [
])
Zm'
=
Ro
+
Zm
=
100
[
] -
j15.9
[
]
Vm
Zm
Vo
Zm
----------------------
15.9
Vo
101.3
--------------------------
=
=
Vo
: generator output
|
Zm'
|
15.9
[
]
×Vo
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......