5.9 Application Settings
222
This function is for outputting a measurement terminal connector error when the measurement result is high
relative to the set judgment reference value. The setting value can be set as an absolute value, and the error
is output via the EXT I/O.
See
"Chapter 11 External Control" (p. 363)
The judgment reference is calculated from the nominal value (range name) of the current measurement range
and the judgment reference value as shown below.
Judgment reference = Nominal value of current measurement range × Judgment reference value (%)
5.9.2 Detecting OPEN during 2-terminal Measurement
(HIGH-Z Reject Function)
Example Current measurement range: 30 k
Judgment reference value: 150%
Judgment reference = 30 k × 1.50 = 45 k
Procedure
Application Settings
1
Analyzer Initial Screen
Press
.
2
Application Settings