5.9 Application Settings
236
•
When the trigger synchronous output function is set to ON, there is a measurement time
delay because the instrument enters a wait time which spans from when the measure-
ment signal is output to when measurement starts.
See
"13.3 About Measurement Times and Measurement Speed" (p. 399)
•
When the trigger synchronous output function is set to ON, the set DC level may be out-
put if a measurement condition is changed.
Also, output will stop when measurement is performed once.
•
The measurement signal is output when the trigger signal is input and stops after mea-
surement ends.
•
In CONTINUOUS measurement mode, the measurement signal stops after measure-
ment of the last panel ends.
Use
or
to set the wait time from after
the measurement signal is output by applying
a trigger to the start of measurement.
4
Settable range: 0.0010 s to 9.9999 s
When you want to return the time to the ini-
tial state:
Press .
The set time is set to 0.0010 s.
Trigger Sync Setting
5
Press
to close the setting screen.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......