11.4 External I/O Settings
379
11
Chapte
r
1
1 External
Co
ntro
l
There are the following setting items for the output timing of the judgment result output signal and the logic of
the trigger signal.
The delay time for the period from the output of the comparator and BIN judgment results until the output of
EOM (LOW) from the EXT I/O can be set on the instrument or by a communication command.
For the setting procedure, refer to the following.
See
Resetting Judgment Results" (p. 115),(p. 226)
Description of communications commands on the included CD (
:IO:OUTPut:DELay
)
Whether to reset the comparator and BIN judgment results simultaneously with the measurement start signal
can be selected on the instrument or by a communication command.
For the setting procedure, refer to the following.
See
Resetting Judgment Results" (p. 115),(p. 226)
Description of communications commands on the included CD (
:IO:RESult:RESet
)
Whether to enable or disable trigger input from the EXT I/O during measurement (during EOM (HI) output)
can be selected on the instrument or by a communication command.
For the setting procedure, refer to the following.
See
Description of communications commands on the included CD (
:IO:TRIGger:ENABle
)
Either the rising edge or falling edge can be selected as the valid edge of trigger input from the EXT I/O.
For the setting procedure, refer to the following.
See
Description of communications commands on the included CD (
:IO:TRIGger:EDGe
)
11.4 External I/O Settings
Setting Delay Time from Output of Comparator and BIN Judgment Results
until Output of EOM (LOW)
Setting Reset of Judgment Results
Enabling Trigger Input for during Measurement
Setting Valid Edge of Trigger Input
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......