4.2 Setting Basic Settings of Measurement Conditions
47
4
Chapte
r 4
LCR Function
• When normal mode (p. 54)
However, the constant current operation range differs depending on the test sample to be measured.
• When Low Z high accuracy mode (p. 54)
However, the constant current operation range differs depending on the test sample to be measured.
Constant current level (CC) setting
Measurement frequency
setting range
Constant current
setting range
Constant current
accuracy
4 Hz to 1.000 MHz
0.01 mA to 50 mA
±10%±10
A
1.0001 MHz to 5.000 MHz
0.01 mA to 10 mA
±20%±10
A
0.010 mA
50.00 mA
10.00 mA
4 Hz
5 MHz
1 MHz
Measurement frequency
setting range
Constant current setting
range
Constant current
accuracy
4 Hz to 100 kHz
0.01 mA to 100.00 mA
±10%±10
A
0.010 mA
100.00 mA
4 Hz
5 MHz
100 kHz
• The constant current that can be set differs depending on the frequency.
• When the test frequency is higher than 1 MHz at an constant current of more than 10 mA, the current is
automatically set to 10 mA.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......