10.5 Loading Measurement Data
348
When a loading error is displayed
Symptom
Check Item, or Cause
Solution/reference
Load Error.(1)
The instrument is not in analyzer mode.
Load measurement data after changing to analyzer
mode.
See "1.3.2 Measurement Mode Selection Screen" (p. 13)
Load Error.(2)
The trigger settings in analyzer mode have
been set to repeat sweep.
Load the measurement data after setting the trigger
settings to sequential sweep or step sweep.
See "5.2.3 Setting the Trigger" (p. 137)
Load Error.(3)
The measurement data file is corrupt.
Load a file that is not corrupt.
Load Error.(4)
There is no measurement data that can be
loaded by the instrument.
Load analyzer data that was saved by the instrument.
Load Error.(5)
The measurement signal setting is not set to
frequency sweep.
Load measurement data after setting the sweep pa-
rameter to frequency sweep.
See "5.2.2 Setting the Sweep Parameter" (p. 135)
Load Error.(6)
Measurement data has been saved in a pa-
rameter format that cannot be loaded.
Verify that the measured values being loaded are the
following parameters:
Z-
, Cs-D, Cs-Rs, Cp-D, Cp-Rp, Ls-Q, Ls-Rs, Lp-Q,
Lp-Rp, Rs-X
Load Error.(7)
The measurement data parameters do not
match the current settings.
Load measurement data after configuring the instru-
ment’s parameters so that they match the parameters
of the measured values being loaded.
See "5.2.1 Setting the measurement parameter" (p. 133)
Load Error.(8)
The measurement data includes erroneous
values.
Example: No measurement after power sup-
ply activation: 828
Overflow: 928
Load measured values that do not contain erroneous
values.
See Description of communications commands on the
included CD (
:MEASure?
)
Load Error.(9)
Parameter information was not saved with the
measurement data.
Load measurement data that includes measurement
parameter information.
See "10.4 Saving Measurement Data" (p. 333)
Load Error.(10)
The number of sweep points in the measure-
ment data does not match the current setting.
Load measurement data after configuring the settings
so that the number of measurement points with which
the instrument has been configured matches the num-
ber of points in the measured values being loaded.
See "5.3.1 Setting Sweep Points" (p. 143)
If an error is displayed after you press
,
you may be experiencing one of the issues listed
below.
Measurement Data Loading Error
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......