13.2 Measurement Range and Accuracy
398
Guaranteed Accuracy Range
The guaranteed accuracy range differs depending on the measurement frequency, measurement signal level,
and measurement range.
Range
DC
4.00 Hz to
99.99 Hz
100.00 Hz to
999.99 Hz
1.0000 kHz to
10.000 kHz
10.001 kHz to
100.00 kHz
100.01 kHz to
1.0000 MHz
1.0001 MHz to
5.0000 MHz
100 M
0.1 V to 2.5 V
0.101 V to 5 V
0.501 V to 5 V
10 M
0.1 V to 2.5 V
0.050 V to 5 V
0.101 V to 5 V
0.501 V to 5 V
1 M
0.050 V to 5 V
0.101 V to 5 V
0.501 V to 1 V
100 k
0.050 V to 5 V
0.101 V to 1 V
30 k
0.05 V to 1 V
10 k
0.005 V to 5 V
3 k
1 k
300
10
1
0.005 V to 5 V
*
2
0.101 V to 5 V
0.501 V to 1 V
100 m
0.1 V to 2.5 V
*
1
0.101 V to 5 V
*
3
0.501 V to 5 V
*
3
The above voltages are the voltage settings equivalent to when V mode.
*1 The guaranteed accuracy is 10 m
or higher
*2 The guaranteed accuracy for when DC bias is 0.101 to 5 V.
*3 The guaranteed accuracy for when DC bias is 10 m
or higher and 1.001 to 5 V.
The above measurement specification was determined using a 1.5C-2 V coaxial cable with
an established cable length for the unit.
Using a cable other than a 1.5C-2 V, or a cable that not an established length for the unit in
question increases the chance of measurement inaccuracy. A large capacitance between
the H terminal and grounding capacitance (GND) or the L terminal and GND may result in
measurement inaccuracy. Please set the GND to 10 pF or less.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......