4.3 Setting DC Resistance Measurement
73
4
Chapte
r 4
LCR Function
This mode balances the L
POT
terminal at 0 V to reduce the influence of contact resistance during low imped-
ance measurement. Furthermore, since the output resistance becomes 10
and sufficient current can flow to
the test sample to be measured, this mode enables highly accurate measurement.
3
Low Z High Accuracy Mode
Procedure
LCR Initial Screen
Rdc Basic Settings
1
2
Press
.
Rdc Basic Settings
3
Range Setting
Select ON/ OFF for the low Z high accuracy
mode.
Sets the low Z high accuracy mode to
OFF.
Sets the low Z high accuracy mode to
ON.
4
Press
to close the setting screen.