7.1 Setting Open Circuit Compensation
279
7
Chap
ter 7 Erro
r Com
pensatio
n
If compensation fails, a window such as the following will be displayed.
When an error message appears and compensation has stopped (when
is touched), open cir-
cuit compensation is turned OFF.
The open circuit compensation process is quite sensitive to noise - both noise originating externally and
induced noise.Therefore, if open circuit compensation has been interrupted with a fault, you should check
the following points before starting the compensation process again:
• Check that the test cables are properly connected.
• Check that nothing is connected to the test cables. (Open circuit compensation cannot be performed
while any test sample is connected to the test cables.)
• Check that the test leads are arranged as closely as possible to their configuration in which measure-
ment will be performed.
• During the compensation process, be sure not to disturb the test cables or to move your hand near
them.
• Execute the guarding process.
See
"Appendix 2 Measurement of High Impedance Components"(p. A3)
Press
in Step
3
of [Open Compensation Setting] (p. 274) to make the compensation data for up
until now invalid and return to the [Initial screen].
When Open Compensation Failed
When You Want to Make Open Circuit Compensation Data Invalid
Error
Open Compensation Setting
The compensation values that are stored internally are not cleared by the operation
described above. When ALL or SPOT is selected, the stored compensation values can be
used.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......