Error Massage and Error Display
Error
overview
Error
Description
Solution and reference
Constant-voltage measurement or constant-
current measurement not possible
This message will be dis-
played to the right of the
monitor value when it is not
possible to perform constant-
voltage measurement or
constant-current measure-
ment.
It will also be displayed if the
voltage applied to the sample
or the current flowing to the
sample exceeds the limit
value (including, for example,
if a current in excess of the
limit value flows to the sample
when the open voltage has
been set to the lowest pos-
sible value).
Reduce the voltage level or
current level.
Less than voltage limit
or current limit
This message will be dis-
played if the set constant
voltage (or constant current) is
not reached because a signal
that is greater than or equal
to the current (or voltage) limit
value is not being applied.
Set the limit value again or
change the measurement sig-
nal level so that the limit value
is not exceeded
"Limit (limiting the voltage and
current applied to the sample)
Load correction fre- quency mismatch
This message will be dis-
played in the settings infor
-
mation if the load correction
frequency does not match the
current measurement fre
-
quency when load correction
is enabled.
Perform load correction at the
same frequency as the mea
-
surement frequency.
Load correction condi
-
tion mismatch
This message will be dis-
played in the settings informa
-
tion if load correction condi
-
tions other than the frequency
do not match the current
measurement conditions when
load correction is enabled.
Perform load correction using
the same conditions as the
measurement conditions.