3
Verifying Package Contents
Probe type (cord length: 1 m)
Open state during open
correction
Shorted state during short
correction
L2001 Pincher Probe
*1
Measurable range: DC to 8 MHz
Maximum applied voltage: DC±30 V
Space between tip electrodes: 0 mm to
approx. 6 mm
Pincer type
Gradations on open
correction fixture
Clamp the tip of the pincers at
the open correction fixture gra
-
dation (using the same value as
the length of the measurement
sample), taking care to insert the
pincers all the way.
(For sample 1005, the length is 1.0
mm.)
Close the tip of the pincers.
Test fixture types
Open state during open
correction
Shorted state during short
correction
9261-10 Test Fixture
Measurable range: DC to 8 MHz
Maximum applied voltage: DC±40 V
Measurement terminal hole diameter:
0.3 mm to 1.5 mm
Cord length: 1 m
Connect the 9261-10 and the
instrument with the connection
cable (do not clamp anything to
the fixture).
Insert the short bar all the way
into the sample mounting area.
9262 Test Fixture
*1
Measurable range: DC to 8 MHz
Maximum applied voltage: DC±40 V
Measurable sample dimensions:
Lead diameter of
φ
0.3 mm to
φ
2 mm
Lead pitch of 5 mm or more
This fixture is for measuring lead com
-
ponents. (less than 10 m
Ω
residual re-
sistance after zero adjustment)
In contact
Turn the knob clockwise to tighten
the sample mounting area.
Insert the short bar all the way
into the sample mounting area.
*1: Although the test fixture appears to use a four-terminal setup, two terminals provide contact with the sample
since H
POT
and H
CUR
as well as L
POT
and L
CUR
are connected inside the fixture and probe.
10
9
8
7
6
5
4
3
2
1
Summary of Contents for IM3536
Page 20: ...16 Operating Precautions ...
Page 34: ...30 Screen Layout and Operation ...
Page 140: ...136 Testing the System Self diagnosis ...
Page 224: ...220 About Measurement Times and Measurement Speed ...
Page 240: ...236 Discarding the Instrument ...