201
Function specifications
Panel save and load
Function
Saves measurement conditions and correction values in the instrument’s in
-
ternal memory and loads saved data. The desired measurement conditions
may be loaded using the following methods:
•Using the keys on the instrument
•
Sending communications commands from an external device to the USB
connector, LAN connector, RS-232C connector, or GP-IB connector on the
rear of the instrument
•
Inputting a signal from an external device to the instrument’s EXT I/O connector
Save types
ALL/HARD/ADJ
•
ALL : Saves all HARD and ADJ content.
•
HARD: Saves measurement conditions and cable length correction setting.
•
ADJ : Saves settings for open correction, short correction, load
correction, and correlation correction and correction values only.
Number of data points that
can be saved
Measurement conditions: Up to 60
Correction values : Up to 128
Default setting
ALL
Beep tone setting
Function
Allows you to set the key tone and judgment result beep tone.
Comparator judgment setting
OFF/ IN/ NG (Default setting: NG)
•
OFF: Does not sound a beep for comparator judgments.
•
IN : Sounds a beep when the result is an IN judgment.
•
NG : Sounds a beep when the result is LO or HI.
Key input setting
OFF/ON (Default setting: ON)
•
OFF: Does not sound a beep when a key is touched.
•
ON : Sounds a beep when a key is touched.
Tone setting
Allows you to select any of four beep tones (A, B, C, or D).
(Default setting: A)
Internal temperature monitoring
Function
•Monitors the temperature inside the instrument.
•
Monitors whether the fan is operating.
•
Displays a warning on the screen in the event of a malfunction.
Error display temperature
threshold
Ambient temperature 50°C
Fan stoppage monitoring
threshold
30% of the rated speed of rotation (4,100 r/min.) or less
USB flash drive operation
Saving measurement data
Function
•
LCR mode:
The SAVE key saves measured values shown on the current screen.
•
Continuous measurement mode:
The SAVE key saves measurement results for the each panel.
Saved data
Measuring instrument information, time and date, measurement conditions,
measurement parameters, and measurement values
Data format
CSV file format
File name
Generated automatically based on the time and date and given the extension “CSV.”