181
External I/O Settings
Setting the delay time (from judgment result output to EOM
―――
output) and
judgment result reset operation
The delay time for the period from the output of the comparator and BIN judgment results until the
output of
EOM
―――
from the EXT I/O can be set.
You can also select whether to reset comparator and BIN judgment results when
EOM
―――
output
changes to HI (indicating that measurement is in progress).
(See "9.2 Example Measurement Timing (Timing Charts)" (p. 171).)
Screen display method (For more information, see p. 24.):
(Measurement screen)
SET
key>(
SET
screen)
ADVANCED
tab>
IO JUDGE
key
1
Set the delay time with the
key.
Settable range: 0.0000 s to 0.9999 s
If you make a mistake, touch the
C
key to
reenter the value.
2
Select the judgment result hold/reset
setting.
OFF
Stores the last judgment results
until the next judgment results are
output.
ON
Resets the judgment results when
the EOM
―――
signal change to HI.
3
Touch the
EXIT
key twice.
Displays the measurement screen.
9
External Control
Summary of Contents for IM3536
Page 20: ...16 Operating Precautions ...
Page 34: ...30 Screen Layout and Operation ...
Page 140: ...136 Testing the System Self diagnosis ...
Page 224: ...220 About Measurement Times and Measurement Speed ...
Page 240: ...236 Discarding the Instrument ...