103
Open Correction
Spot correction
Acquire the correction values at the set measurement frequencies. Measurement frequencies can
be set for up to five points.
Screen display method (For more information, see p. 27):
(Measurement screen)
ADJ
key>(
ADJ
screen)
OPEN
key
1
Touch the
ADJUST
key.
2
Touch the
SPOT
key and then touch the
EXIT
key.
1
2
3
Select the correction point you want to
set or edit with the
key, and touch
the
EDIT
key.
2
1
Touch when you wish to cancel correction.
(The instrument will return to the screen
shown in step 2.)
4
Enter the frequency to correct with the
numeric keypad and touch the
SET
key
to accept the setting.
1
2
The previous value will be displayed until
you enter a value.
•
Settable range: DC, 4 Hz to 8 MHz
*
*: The maximum frequency varies with the
• Touch the
C
key to cancel input.
•
If you attempt to set a measurement fre
-
quency greater than the maximum fre
-
quency for each cable length setting, it will
automatically be reduced to the maximum
frequency for each cable length setting.
•
If you attempt to set a measurement fre
-
quency lower than 4 Hz, it will automatically
be increased to 4 Hz.
The correction values from last time are
displayed in a confirmation screen.
Check that the measurement cable is in an
open circuit state.
5
Error Correction