Measurement Range and Accuracy
No.
Frequency band
Measurement speed coefficient
16
8
4
2
1
19 300.01 kHz to 400.00 kHz
1
2 to 19
20 to 79
80 to 1599
1600
20 400.01 kHz to 500.00 kHz
1
2 to 19
20 to 79
80 to 1599
1600
21 500.01 kHz to 700.00 MHz
1 to 3
4 to 31
32 to 159
160 to 2399
2400
22 700.01 kHz to 1.0000 MHz
1 to 3
4 to 31
32 to 159
160 to 2399
2400
23 1.0001 MHz to 1.4000 MHz
1
2 to 13
14 to 63
64 to 959
960
24 1.4001 MHz to 2.0000 MHz
1
2 to 13
14 to 63
64 to 959
960
25 2.0001 MHz to 3.0000 MHz
1 to 2
3 to 23
24 to 95
96 to 1439
1440
26 3.0001 MHz to 4.0000 MHz
1 to 3
4 to 39
40 to 159
160 to 2399
2400
27 4.0001 MHz to 5.0000 MHz
1 to 3
4 to 39
40 to 159
160 to 2399
2400
28 5.0001 MHz to 6.0000 MHz
1 to 7
8 to 63
64 to 249
250 to 3999
4000
29 6.0001 MHz to 8.0000 MHz
1 to 7
8 to 63
64 to 249
250 to 3999
4000
E: Coefficient of the measurement cable's length
The coefficient of the measurement cable's length is shown in the table below.
Measurement cable's length
0 m
1 m
2 m
4 m
Coefficient
1
1.5
2
3
The settable range for frequency varies with the measurement cable’s length.
Cable length
Settable range for frequency
0 m
Up to 8 MHz
1 m
Up to 8 MHz
2 m
Up to 2 MHz
4 m
Up to 1 MHz
F: DC bias coefficient
The DC bias coefficient is shown in the table below.
DC bias coefficient
OFF
ON
Coefficient
1
2
G: Temperature coefficient
The operating temperature coefficient is shown in the table below.
(When the operating temperature (t) is 23°C±5°C, use a coefficient of 1.)
Operating temperature
t [°C]
Coefficient
1+0.1×|t-23|
10
Specifications