Saving Measurement Conditions and Correction Values (Panel Save Function)
6.1 Saving Measurement Conditions and Correction
Values (Panel Save Function)
This section describes how to save measurement condition data and correction value data in the
instrument’s memory. The following number of data sets can be saved:
(Measurement condition: Up to 60 items, Correction value: Up to 128 items)
First, select the data type you wish to save. You can select from three types (see procedure below).
Then, save the selected type of data to a panel (See p. 126).
Setting the type of data to save
Screen display method (For more information, see p. 24.)
(Measurement screen)
SET
key>(
SET
screen)
ADVANCED
tab>
PANEL
key
1
Touch the
SAVE TYPE
key.
2
Select the type of data to save and
touch the
EXIT
key.
1
2
ALL
Saves all
HARD
and
ADJ
content.
(Indication on screen:
LCR+ADJ
)
HARD
Saves the measurement condition
and cable length correction setting
value. (Indication on screen:
LCR
)
ADJ
Saves only each of the setting values
and correction values of open correc
-
tion, short correction, load correction,
and correlation (scaling) correction.
(Indication on screen:
ADJ
)
3
Touch the
EXIT
key twice.
Displays the measurement screen.
When the type of data to save is set to
ALL
, data is saved to one panel but counted as one set of measure
-
ment conditions and one set of correction values.
(Example: When data is saved after setting the type of data to save to
ALL
, the panel is counted as 1
LCR
[measurement conditions] data set and 1
ADJ
[correction values] data set.)
Each count will increase by 1.