166
External Input/Output Connector and Signals
Pin
I/O
*
Signal name
Function
Logic
19
-
(Unused)
-
-
-
20
-
(Unused)
-
-
-
21
-
(Unused)
-
-
-
22
-
(Unused)
-
-
-
23
-
(Unused)
-
-
-
24
-
(Unused)
-
-
-
25
-
(Unused)
-
-
-
26
-
(Unused)
-
-
-
27
-
ISO_COM
Isolated common signal ground
-
-
28
OUT EOM
―――
This signal indicates that measurement is complete. “HI
(OFF)” indicates that measurement is in progress, while
“LO (ON)” indicates that measurement is complete. When
LO (ON), the comparator judgment results have been
finalized.
Falling
Edge
29
OUT
INDEX
―
―――
―
Signal indicating that A/D conversion for the measurement
circuit has completed: When this signal changes from HI
(OFF) to LO (ON), the sample may be changed.
Falling
Edge
30
OUT
No.1_PARA1-IN
―
―― ―
―――
――
―
Outputs IN comparator judgment results for the first No. 1
parameter.
Negative
Level
31
OUT
No.1_PARA3-HI
―
―― ―
―――
――
―
Outputs HI comparator judgment results for the first No. 3
parameter.
Negative
Level
32
OUT
No.1_PARA3-LO
―
―― ―
―――
――
―
Outputs LO comparator judgment results for the first No. 3
parameter.
Negative
Level
33
OUT
No.2_PARA1-HI
―
―― ―
―――
――
―
Outputs HI comparator judgment results for the second
No. 1 parameter.
Negative
Level
34
OUT
No.2_PARA1-LO
―
―― ―
―――
――
―
Outputs LO comparator judgment results for the second
No. 1 parameter.
Negative
Level
35
OUT
No.2_PARA3-IN
―
―― ―
―――
――
―
Outputs IN comparator judgment results for the second
No. 3 parameter.
Negative
Level
36
-
(Unused)
-
-
-
37
-
(Unused)
-
-
-
*: IN indicates signal input to the instrument, while OUT indicates signal output from the instrument.