If Open or Short Correction Fails to Complete Normally
(2) A window such as the following will be displayed if the cable length setting does not
match the length of the connected cable (during open correction only).
A window such as the following will be displayed.
Adjusting
To change the cable length setting, touch
the
CANCEL
key.
Completed
Touching the
EXIT
key will enable the ac-
quired correction value.
Solution
•
Verify that the length of the connected cable and the cable length setting mach (p. 98).
•
The connected cable length is detected based on the voltage monitor value. It may not be
possible to properly detect the cable length depending on the cable type and length and the
impedance value at the time of open correction.