106
Short Correction
All correction
Simultaneously acquire the short correction values for all measurement frequencies.
Screen display method (For more information, see p. 27):
(Measurement screen)
ADJ
key>(
ADJ
screen)
SHORT
key
1
Touch the
ADJUST
key.
2
Touch the
ALL
key, and then touch the
EXIT
key.
1
2
The correction values from last time are
displayed in a confirmation screen.
(If correction has never been performed,
the correction values become 0.)
Check that the measurement cable is in an
short circuit state.
3
Touch the
EXEC
key.
Touch when you wish to
cancel correction.(The dis-
play will return to the screen
shown in step 2, and the
short correction value will
remain unchanged.)
The correction
range can be
limited. (p. 101)
Correction starts.
Compensation value acquisition time:
Approx. 50 seconds
Touch when you wish to cancel correction.
(The display will return to the screen shown
in step 2, and the short correction value will
remain unchanged.)
The next screen will be displayed once cor-
rection has completed normally.
Correction results
(Effective resistance, reactance)
Correction
No.
Measurement frequency
•
You can check the effective resistance
and reactance at each correction point
with the
keys.
• The possible correction range is 1 k
Ω
or
less for impedance. Correction cannot be
performed if the measured value (cable
or fixture residual impedance) is 1 k
Ω
or
greater.
4
Touch the
EXIT
key twice.
Displays the measurement screen.
•
If correction does not complete normally:
•