162
External Input/Output Connector and Signals
9.1 External Input/Output Connector and Signals
Before connecting the terminals, be sure to read "Before Connecting EXT I/O" (p. 15).
This section describes the instrument’s EXT I/O connectors, compatible connectors, connector signal
assignments, input (IN) signal functionality, and output signals when errors occur.
Signal input or output is indicated as “LO (ON),” while the absence of signal input or output is indicated as “HI
(OFF).” (Please note that this usage differs in meaning from “HI” and “LO” as used for judgment results.)
Instrument connector and supported connectors
The instrument provides the following EXT I/O connector and supports use of the following types of
connectors:
Instrument side connector
37-pin D-sub female with #4-40 screws
Mating connectors
•DC-37P-ULR (solder type)
•
DCSP-JB37PR (pressure weld type)
Japan Aviation Electronics Industry Ltd.
Instrument connector signal assignments
Signal assignments vary with the measurement mode.
Signal logic is 0 V to 0.9 V for LO level and 5 V to 24 V for HI level.
LCR mode (LCR) operation
2 1
4 3
6 5
8 7
9
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
14
15
16
17
18
19
10
11
12
13
TRIG
―
――
―
(Unused)
(Unused)
LD1
―
―
―
LD3
―
―
―
LD5
―
―
―
C1
―
―
ISO_5V
ISO_COM
ERR
――
―
BIN1
―――
―
, P
ARA1-HI
―
―――
―
―
―
, D1-0
―
―
―
BIN3
―――
―
, P
ARA1-LO
―
―――
―
―
―
, D1-2
―
―
―
BIN5
―――
―
, P
ARA3-IN
―
―――
―
―
―
, D2-0
―
―
―
BIN7
―――
―
, AND
―――
, D2-2
―
―
―
BIN9
―――
―
, D3-0
―
―
―
D3-2
―
―
―
D4-0
―
―
―
D4-2
―
―
―
OUT_OF_BINS
―
―
―
―
―
―――
―
C2
―
―
CALIB
―
―
――
―
LD0
―
―
―
LD2
―
―
―
LD4
―
―
―
LD6
―
―
―
LD_V
ALID
―
―
―
―
―
―
―
ISO_COM
EOM
――
―
INDEX
――
―
―
―
BIN2
―――
―
, P
ARA1-IN
―
―――
―
―
―
D1-1
―
―
―
BIN4
―――
―
, P
ARA3-HI
―
―――
―
―
―
, D1-3
―
―
―
BIN6
―――
―
, P
ARA3-LO
―
―――
―
―
―
, D2-1
―
―
―
BIN8
―――
―
, D2-3
―
―
―
BIN10
―――
―
―
, D3-1
―
―
―
D3-3
―
―
―
D4-1
―
―
―
D4-3
―
―
―
Summary of Contents for IM3536
Page 20: ...16 Operating Precautions ...
Page 34: ...30 Screen Layout and Operation ...
Page 140: ...136 Testing the System Self diagnosis ...
Page 224: ...220 About Measurement Times and Measurement Speed ...
Page 240: ...236 Discarding the Instrument ...