200
Function specifications
• Spot correction
Function
Acquires correction values for the set measurement frequencies.
Maximum number of
settings
5 points
Valid frequency setting
range
DC, 4 Hz to 8 MHz
•
OFF (Default setting)
Function
Disables short correction data.
Load correction
Function
Measures a reference sample whose measured value is known and then
calculates the correction coefficient. The coefficient is used to correct mea
-
sured values.
Number of sets of
correction conditions
Up to 5
Correction correction
settings
Correction frequency, correction range, correction signal level, DC bias,
parameter type, reference value 1, reference value 2
Seven settings can be configured for each set of correction conditions.
Calculation equation
Z
=(
Z
before correction )×(
Z
correction coefficient)
θ
=(
θ
before corr(
θ
correction value)
Z
correction coefficient=
(
Z
reference value)
(accutual data of
Z
)
θ
correction value =(
θ
reference value)-(accutual data of
θ
)
Operation mode
OFF/ON
Default setting
OFF
Cable length correction
Function
Corrects measurement errors caused by the effects of the measurement
cable.
Cable length settings
0 m, 1 m, 2 m, 4 m
Default setting
0 m
Correlation correction
Function
•
Corrects measured values using user-configured correction coefficients
(using user-configured correction coefficient A and correction coefficient B).
•
Corrected measured values are calculated using the following equation:
(Measured value after correction)=A×(measured value)+B
Operation mode
OFF/ON
Correction coefficient A
setting range
-999.999 to 999.999
Correction coefficient B
setting range
-9.99999 G to 9.99999 G
Default setting
OFF
(Correction coefficient A: 1.000, Correction coefficient B: 0.00000)