230
Error Massage and Error Display
11.3 Error Massage and Error Display
If a message or display such as those shown below is displayed on the screen, refer to the indi
-
cated section for more information.
Error
overview
Error
Description
Solution and reference
Backup battery low
The life of the RAM backup
battery has ended.
The instrument needs to be
repaired.
Contact your authorized Hioki
distributor or reseller.
Memory full
This message will be shown
on the top of the screen if the
set number of measurement
results has been stored in the
instrument’s internal memory.
Once this message has been
displayed, you will not be able
to save any more measured
values.
Load measurement values
stored in the instrument’s
memory with the memory
function or clear the memory.
No measured value ac-
curacy guarantee
This message will be shown
on the top of the screen if the
accuracy of the measured
value is not guaranteed.
Increase the measurement
signal level or change the
measurement range to one
that matches the impedance
of the element to be mea
-
sured.
Hi Z reject error
This message will be shown
on the top of the screen if
the measurement results are
higher than the judgment
standard that was set with the
Hi Z reject function.
•
Check the connection of
each terminal.
•
Check the Hi Z reject func
-
tion settings.