213
Measurement Range and Accuracy
The range of measurement levels for which accuracy is guaranteed varies with the setting condi
-
tions.
Guaranteed accuracy measurement level range
Range
Sample's
imped-
ance
Measurement frequency
DC
4 Hz to
99.99 Hz
100 Hz to
999.99 Hz
1 kHz to
10 kHz
10.001 kHz
to 100 kHz
100.01 kHz
to 1 MHz
1.0001 MHz
to 5 MHz
5.0001 MHz
to 8 MHz
100 M
Ω
8 M
Ω
to
200 M
Ω
1 V (fixed)
0.101 V to 5 V
10 M
Ω
10 M
Ω
to
100 M
Ω
0.101 V to 5 V
800 k
Ω
to
10 M
Ω
0.101 V to 5 V
0.501 V
to 5 V
1 M
Ω
1 M
Ω
to
10 M
Ω
0.101 V to 5 V
0.501 V
to 5 V
80 k
Ω
to
1 M
Ω
0.050 V to 5 V
0.101 V
to 5 V
0.501 V
to 5 V
100 k
Ω
100 k
Ω
to
1 M
Ω
0.050 V to 5 V
0.101 V
to 5 V
0.501 V
to 5 V
8 k
Ω
to
100 k
Ω
0.010 V to 5 V
0.050 V
to 5 V
0.101 V
to 1 V
10 k
Ω
10 k
Ω
to
100 k
Ω
0.010 V to 5 V
0.050 V
to 5 V
0.101 V
to 1 V
800
Ω
to
10 k
Ω
0.010 V to 5 V
0.050 V
to 5 V
0.050 V
to 1 V
0.101 V
to 1 V
1 k
Ω
1 k
Ω
to
10 k
Ω
0.010 V to 5 V
0.050 V
to 5 V
0.050 V
to 1 V
0.101 V
to 1 V
80
Ω
to
1 k
Ω
0.010 V to 5 V
0.050 V
to 1 V
0.101 V
to 1 V
100
Ω
8
Ω
to
100
Ω
0.010 V to 5 V
0.050 V
to 1 V
0.101 V
to 1 V
10
Ω
800 m
Ω
to 10
Ω
0.050 V to 5 V
0.101 V to 1 V
1
Ω
80 m
Ω
to
1
Ω
0.050 V to 5 V
0.101 V
to 5 V
0.501 V to 1 V
100 m
Ω
1 m
Ω
to
100 m
Ω
0.101 V to 5 V
0.501 V
to 5 V
The accuracy guarantee range during DC bias operation is 10 m
Ω
or greater.
The accuracy for DC resistance (Rdc) measurement is guaranteed only when offset values are
acquired.
The guaranteed accuracy range varies with the sample’s impedance.
10
Specifications
Summary of Contents for IM3536
Page 20: ...16 Operating Precautions ...
Page 34: ...30 Screen Layout and Operation ...
Page 140: ...136 Testing the System Self diagnosis ...
Page 224: ...220 About Measurement Times and Measurement Speed ...
Page 240: ...236 Discarding the Instrument ...