208
Measurement Range and Accuracy
Range
For AC measurement (measurement frequency)
10.001 kHz to 100.00 kHz
100.01 kHz to 1 MHz
1.0001 MHz to 8 MHz
100 M
Ω
-
-
-
-
-
-
-
-
-
-
-
-
10 M
Ω
A=2
B=1
-
-
-
-
A=2
B=1
-
-
-
-
1 M
Ω
A=0.5
B=0.1
A=3
B=0.5
-
-
A=0.6
B=0.1
A=3
B=0.5
-
-
100 k
Ω
A=0.25
B=0.04
A=1
B=0.3
A=2
B=0.5
A=0.2
B=0.02
A=1
B=0.3
A=2
B=0.3
10 k
Ω
A=0.3
B=0.02
A=0.5
B=0.05
A=2
B=0.5
A=0.2
B=0.02
A=0.5
B=0.05
A=1.5
B=0.3
1 k
Ω
A=0.2
B=0.02
A=0.4
B=0.02
A=1.5
B=0.2
A=0.15
B=0.02
A=0.4
B=0.02
A=1.5
B=0.2
100
Ω
A=0.2
B=0.02
A=0.5
B=0.03
A=1.5
B=0.2
A=0.15
B=0.02
A=0.5
B=0.03
A=1.5
B=0.2
10
Ω
A=0.4
B=0.05
A=0.8
B=0.1
A=2
B=1.5
A=0.3
B=0.03
A=0.5
B=0.05
A=2
B=1
1
Ω
A=1
B=0.3
A=1.5
B=1
A=3
B=3
A=0.5
B=0.2
A=0.7
B=0.5
A=3
B=2
100 m
Ω
A=2
B=2
A=4
B=3
-
-
A=2
B=1.5
A=3
B=4
-
-
Summary of Contents for IM3536
Page 20: ...16 Operating Precautions ...
Page 34: ...30 Screen Layout and Operation ...
Page 140: ...136 Testing the System Self diagnosis ...
Page 224: ...220 About Measurement Times and Measurement Speed ...
Page 240: ...236 Discarding the Instrument ...