217
About Measurement Times and Measurement Speed
10.7 About Measurement Times and Measurement
Speed
Measurement times vary with the measurement conditions (see the table below). Values are all
provided for reference purposes only (and vary with operating conditions).
Analog measurement signal (INDEX
―
――――
)
Measurement speed
Measurement
frequency
FAST
MED
SLOW
SLOW2
DC (line frequency 50 Hz)
1 ms
20 ms
100 ms
400 ms
DC (line frequency 60 Hz)
1 ms
16.67 ms
100 ms
400 ms
4.00 Hz to 10.00 Hz
Tf
2×Tf
3×Tf
4×Tf
10.01 Hz to 39.99 Hz
Tf
2×Tf
5×Tf
10×Tf
40.00 Hz to 99.99 Hz
Tf
2×Tf
5×Tf
40×Tf
100.00 Hz to 300.00 Hz
Tf
2×Tf
5×Tf
50×Tf
300.01 Hz to 500.00 Hz
Tf
2×Tf
10×Tf
200×Tf
500.01 Hz to 1.0000 kHz
Tf
5×Tf
20×Tf
300×Tf
1.0001 kHz to 2.0000 kHz
Tf
8×Tf
40×Tf
600×Tf
2.0001 kHz to 3.0000 kHz
2×Tf
12×Tf
60×Tf
1200×Tf
3.0001 kHz to 5.0000 kHz
3×Tf
20×Tf
100×Tf
2000×Tf
5.0001 kHz to 10.000 kHz
5×Tf
40×Tf
200×Tf
3000×Tf
10.001 kHz to 20.000 kHz
10×Tf
80×Tf
400×Tf
6000×Tf
20.001 kHz to 30.000 kHz
25×Tf
150×Tf
600×Tf
12000×Tf
30.001 kHz to 50.000 kHz
25×Tf
250×Tf
1000×Tf
20000×Tf
50.001 kHz to 100.00 kHz
50×Tf
400×Tf
2000×Tf
30000×Tf
100.01 kHz to 140.00 kHz
100×Tf
800×Tf
4000×Tf
60000×Tf
140.01 kHz to 200.00 kHz
100×Tf
800×Tf
4000×Tf
60000×Tf
200.01 kHz to 300.00 kHz
250×Tf
1500×Tf
6000×Tf
120000×Tf
300.01 kHz to 400.00 kHz
250×Tf
2500×Tf
10000×Tf
200000×Tf
400.01 kHz to 500.00 kHz
250×Tf
2500×Tf
10000×Tf
200000×Tf
500.01 kHz to 700.00 kHz
500×Tf
4000×Tf
20000×Tf
300000×Tf
700.01 kHz to 1.0000 MHz
500×Tf
4000×Tf
20000×Tf
300000×Tf
1.0001 MHz to 1.4000 MHz
1250×Tf
8750×Tf
40000×Tf
600000×Tf
1.4001 MHz to 2.0000 MHz
1250×Tf
8750×Tf
40000×Tf
600000×Tf
2.0001 MHz to 3.0000 MHz
1875×Tf
15000×Tf
60000×Tf
900000×Tf
3.0001 MHz to 4.0000 MHz
2500×Tf
25000×Tf
100000×Tf
1500000×Tf
4.0001 MHz to 5.0000 MHz
2500×Tf
25000×Tf
100000×Tf
1500000×Tf
5.0001 MHz to 6.0000 MHz
5000×Tf
40000×Tf
156250×Tf
2500000×Tf
6.0001 MHz to 8.0000 MHz
5000×Tf
40000×Tf
156250×Tf
2500000×Tf
•
Tf [sec]: 1/measurement frequency [Hz]
•
The above DC measurement times apply to operation with DC adjustment off. To calculate times for opera
-
tion with DC adjustment on, multiple the above times by 2 and add 0.5 ms.
10
Specifications
Summary of Contents for IM3536
Page 20: ...16 Operating Precautions ...
Page 34: ...30 Screen Layout and Operation ...
Page 140: ...136 Testing the System Self diagnosis ...
Page 224: ...220 About Measurement Times and Measurement Speed ...
Page 240: ...236 Discarding the Instrument ...