111
Disabling Open and Short Correction Values
5.5 Disabling Open and Short Correction Values
Turning off the correction setting will disable the correction values you have acquired.
Screen display method (For more information, see p. 27):
To disable open correction: (Measurement screen)
ADJ
key>(
ADJ
screen)
OPEN
key
To disable short correction: (Measurement screen)
ADJ
key>(
ADJ
screen)
SHORT
key
1
Touch the
ADJUST
key.
2
Touch the
OFF
key and then touch the
EXIT
key.
1
2
3
Touch the
EXIT
key twice.
Displays the measurement screen.
The correction values that are stored internally are not cleared by the operation described above. When
ALL
or
SPOT
is selected, the stored correction values can be used.