43
Setting Measurement Conditions (basic settings)
3.4 Setting Measurement Conditions (basic set-
tings)
(
There are two types of measurement: AC measurement and DC measurement (p. 40).
The measurement conditions set for AC measurement and DC measurement differ.
Required: Be sure to set.
Optional: Change setting as necessary.
Setting
During AC
measurement
(When the parameter
is other than
Rdc
)
During DC
measurement
(When the parameter
is other than
Rdc
)
Ref.
Overview
Measurement
frequency
Required
-
Configure based on the mea
-
surement sample.
Measurement range
Required
Required
Measurement signal
level
Required
-
Line frequency
-
Required
p. 54 Set to the power supply’s fre
-
quency.
Measurement speed
Optional
Optional
When you wish to perform mea
-
surement more quickly:
FAST
When you wish to perform mea
-
surement at a higher level of
precision:
SLOW
, or
SLOW2
Low Z high accuracy
mode
Optional
(default value:
OFF
)
Optional
(default value:
OFF
)
To perform high-precision mea
-
surement, set to
ON
.
To measure at high speed:
OFF
Average
Optional
(default value:
OFF
)
Optional
(default value:
OFF
)
p. 57 To limit variability of displayed
values, set to
ON
.
Limit
Optional
(default value:
OFF
)
-
To limit the voltage or current
that is applied to the sample, set
to
ON
.
DC bias
Optional
(default value:
OFF
)
-
To superimpose a DC voltage on
the measurement signal during
measurement, set to
ON
.
DC adjust
-
Optional
(default value:
OFF
)
To reduce measurement error:
ON
To measure at high speed:
OFF
Trigger Synchronous
Output
Trigger Synchronous
delay
*
Optional
(default value:
OFF
, default value: 0.0010 s)
To apply the signal during mea-
surement only, set to
ON
.
Optional (default value: 0.0010 s)
Set to a sufficiently large value
when you wish to stabilize mea-
surement.
DC delay
*
-
Optional
(default value: 0 s)
Adjustment delay
*
-
Optional
(default value: 0.0030 s) p. 63
Trigger
Optional (Default setting:
INT
)
Measurement is repeated automatically.
To input signals and commands
from an external source, set to
EXT
.
Trigger delay
*
Optional (default value: 0 s)
If the trigger function is enabled,
set to a large enough value that
measurement can stabilize.
3
Performing Measurements in LCR Mode