REMOTE SETPOINT MENU
PERCENTAGE % 0. 0
LO 0. 000 LBS/M IN
SPH 3.000 LBS/MIN
Chapter 8 - Operation
8-21
FIG. 8-24 SETTING LO AND HI SIGNAL OUTPUT PARAMETERS
Step 4. Setting the HI Output Signal parameter
•
The HI setting is the maximum signal output to be refer-
enced as 3.000 lbs/min (Default Setting). Depending on
the output option board purchased the output signal range
could be 0-20 milliamps, 4-20 milliamps, 0-5 volts, 0-10
volts. The HI setting corresponds to 20 milliamp or 5 or
10 volts output.
•
The Default setting is 3.000 mass units/time units. The
value must be higher than the LO value setting.
•
Press the down arrow until the cursor is to the left of the
HI setting sub-menu.
•
Press the Clear button to clear the old value.
NOTE:
If the old value is Zero, enter a new number. If the old value is not Zero, press the
Clear button two times, then enter the new number.
•
Use the numeric keys to enter the new value.
•
Press the Enter button to accept the new value.
•
Press the down arrow until the cursor is to the left of the
Averages sub-menu.
Step 5. Setting the Averaging parameter (See Fig. 8-25)