Unable to detect HR
ECG cable and leads are connected to monitor and patient well, but HR is
unable to be detected. It may be caused by poor integrity of ECG signals.
Unable to detect SpO
2
SpO
2
probe is connected to monitor and patient well, but SpO
2
is unable
to be detected. It may be caused by poor plethysmogram
signals.
Low battery
Low battery voltage
ECG Lead Off
The ECG electrodes or cable fell off
SpO
2
Probe Off
SpO
2
probe fell off
B - Status/Error during Monitoring
“Cuff error”
cuff is not wrapped correctly, or is not connected
“Air leak”
Air moving part, tube or the cuff leak air.
“Pressure error”
Unstable cuff pressure or tangled cuff tubing
“Signal weak”
Very weak signal because of the cuff, or the patient has very weak pulse
“Over extent”
The measurement range exceeds 255 mmHg (Pediatric patient over 135 mmHg)
“Over motion”
The repeated measurement due to moving, excessive noise during the stepping inflation and
measuring pressure and pulse, e.g. during patient shaking motion
“Signal overflow”
Blood pressure amplifier overflow due to excessive movement
“Leak in gas run”
Leaking during the pneumatic device testing
“System error”
Abnormal condition of CPU, such as register overflow, divided by zero
“Adult”
The blood pressure measuring now is in adult mode. In this case, it is not allowed to
monitoring Pediatric or neonatal patient. Otherwise, there may be serious danger to the
Pediatric monitored.
“Pediatric”
The blood pressure module is now worked in Pediatric measuring mode.
“PROBE OFF”
SpO
2
probe fell off
“LEADS OFF”
The ECG electrodes or cable fell off
“LEARNING”
Learning arrhythmia for 15 seconds
“DEMO”
The monitor is displaying the demo waveforms, which are generated by the monitor itself.