47
ENGLISH
6.3.1 Alarm lamp
Lamp color
Alarm level
Red flashing
High priority alarm
Yellow flashing
Medium priority alarm
Yellow
Low priority alarm
Green
Normal
6.3.2 Alarm message
When an alarm occurs, an alarm message will appear in the technical or physiological alarm area. And the alarm message
is shown in different color to match the alarm level (high priority alarm in red, medium priority alarm in yellow, and low
priority alarm in white).
The alarm log will be stored permanently in the monitor, even if the power is down accidental or total loss of power, but
the event of accidental power down will down will not capture in the log.
The monitor can store alarm events, a single patient maximum of 1000 groups, when the number of events reach the
maximum storage capacity, the latest event will cover the earliest historical events, that is, the event displayed is the
current patient’s most recent 1000 events.
6.3.3 Highlighting Numeric
If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in alarm will be highlighted.
6.3.4 Audible Alarm Tones
The alarm tone is distinct from the heart beat tone, keystroke tone and pulse tone in frequency, the alarm tones identify
the alarm levels as follows:
High priority alarm: beep+beep+beep+pause+beep+beep+beep
Medium priority alarm: triple beep
Low priority alarm: single beep
When multiple alarms of different levels occur simultaneously, the monitor will select the alarm of the highest
level and give visual and audible alarm indications accordingly.
6.3.5 Alarm Status Symbols:
Apart from the aforementioned alarm indicators, the monitor still uses the following symbols telling the alarm status.
For the customer purchased the monitor with configuration of “
Alarm pause
”, then this area shows the alarm sound
status. There are 3 alarm sound status:
①”
” means the alarm sound is enabled;
② “
” means the alarm sound is disabled or means
the alarm volume is 0 (alarm is mute), at this time the user should
pay more attention to the patient
;
③ “
” means the alarm sound is paused. Text display area: when the alarm sound is paused, it displays