43
ENGLISH
5.5 NIBP Related Settings
On “System Menu” screen, select “Setup”
→
“NIBP” to enter into NIBP related settings.
Figure 5.5 NIBP related settings
Unit
: The pressure unit, and mmHg and kPa can be selected. The factory default is mmHg.
Mode
:
The measuring mode, manual or automatic. The factory default is manual. The operator needs to press the
NIBP button to perform blood pressure measurement. If the Auto mode is chosen, the operator needs to set an cycle
interval as well.
Cycle:
The time interval between measurements when the measuring mode is set to Auto. The options are STAT, 1
min, 2 min…480 min. Press NIBP to start measurement and the monitor begins to count down as well. It will take
blood pressure measurement automatically after finishing counting down. If “STAT” is selected, press NIBP to take
the blood pressure measurement as many times as possible, but it will sustain this state not longer than 5 minutes.
WARNING: STAT can only be used for Adult. Using this mode to Pediatric patient can cause
serious injury.
NIBP Cali (BP verification):
It is used to check the accuracy of pressure measuring system within NIBP module,
which should be conducted by technicians in test house or service department. There are three options: “Mode 1”,
“Mode 2” and “OFF” available. After the verification, it is necessary to make sure it is in “OFF” status again, or other
operations can’t be carried out and NIBP key will be inactivated. Factory default status is “OFF”.
Gas Leak:
it is used by technicians to perform a leakage inspection for NIBP pneumatic system.
SYS Hi/Lo
:
High and Low alarm limits for systolic pressure
DIA Hi/Lo
: High and Low alarm limits for diastolic pressure
MAP Hi/Lo
: High and Low alarm limits for MAP
PR Hi/Lo
: High and Low alarm limits for PR
Initial inflation pressure setting:
Cuff pressure to be inflated initially, its options are different
depending on patient type.
for neonate:
initial inflation pressure can be: 60, 70, 80mmHg, default setting: 70 mmHg;
for Pediatric:
initial inflation pressure can be: 80, 100, 120, 140 mmHg, default setting: 100 mmHg;
for adult:
initial inflation pressure can be: 80, 100, 120, 140, 160, 180, 200mmHg, default setting: 150
mmHg.