![Gima 35134 Operator'S Manual Download Page 208](http://html1.mh-extra.com/html/gima/35134/35134_operators-manual_2228002208.webp)
20
ENGLISH
2.3.1 Information Area
The Information Area will display the patient information (including gender, patient ID, patient type and name),
physiological alarm sources, alarm sound status and current date.
Alarm status Patient type ECG Filter type Network connection indicator Date and Time
“Alarm
”:
Audible alarm status, green “
” indicates the audible alarm is enabled, yellow “
” indicates
the alarm sound is silent. The audible alarm will be activated again automatically after 2 minutes or when a new
alarm event occurs. Red “ ” indicates the audible alarm is disabled, that means the alarm sound is off, this is also
normal situation when the alarm sound volume is set to “0” in system parameter settings.
“ADUL”:
The patient type. There are two types available: “Adult” and “Pediatric”
.
“MON”:
ECG Filter type. There are three types: “Diagnosis”, “Monitor” and “Operation”. The option can be set in
the setting menu.
: network connection indicator. “
” means the patient monitor is not linked with the central monitoring
system yet; “
” (in green) means the patient monitor is linked to the central monitoring system successfully.
:
battery voltage indicator; When the indicator is yellow and displays only one “grid”, it means there is a
little battery power left. When the indicator turns red and blinks, as well as less than one “grid” displays, the system
alarm will be on to remind the low battery voltage. Please connect the device to the mains power supply in time
to ensure the normal use of monitor, and the battery will be recharged. When the battery is fully charged, battery
voltage indicator displays full grid. During recharging, the grids in the battery indicator are rolling circularly.
“2011-03-02 13:57:25”:
the current calendar time and date. The system time and date can be set during the system
start-up when the screen displays the time and data setups. The current figure shows the time and date is March
2
nd
, 13:57:25, 2011.
“Long press knob to enter System Menu”:
System prompt or description for the current status.
“PID”:
The patient ID. The patient ID can be entered or changed in the archive management window