Chapter 5 Parameter Settings
Step 1:
rotate the knob to move the gray cursor to the setting item and press the knob to confirm your selection.
Step 2:
rotate the knob to change the setting or modify the setting value.
Step 3:
press the knob again to change and repress it to save the setting.
Pressing “
” key to return to the Main Screen.
Long pressing Navigation Knob to enter into “System Menu” screen.
5.1 System Parameter Settings
On “System Menu” screen, select “Setup”
→
“SYS” to enter into system related settings.
Figure 5.1 System Setup
Type
: The category of patient being monitored, this can be selected from Adult, Pediatric and Neonate.
Adult
: the patient is adult.
Pediatric
: the patient is pediatric.
Neonate:
the patient is neonate.
The default is “Adult”
When changing the patient type, the monitor will change the default alarm settings, and initialize the NIBP
and SpO
2
measuring module accordingly. Please pay special attention to the patient type before starting the
monitoring. It is forbidden to use Adult type on the Pediatric patient, or it may cause serious injury.
Mode
: Monitor working mode selection. The “Real Time” shows the normal working mode with real time waveform
and data. The “Demo” shows the demo working mode with demo waveform and data. In the demo mode, all the
signals and data are generated by the device for demonstrating and testing purpose. The default is “Real Time”
LANG
:
The current language used, which can be selected by the user. There is no default for this setting. However,
the setting can be saved.
Fill
: When the fill setting is ON, the plethysmogram and respiration waveform will be displayed with filled curves.
When it is OFF, the monitor displays the line graph. The default is OFF.
Frze
:
Pressed the key to freeze the selected waveforms. The options are “All” and “ECG”. When “ECG” is selected,
the monitor only freezes the ECG waveform. When “All” is selected, the monitor freezes all the waveforms including
ECG, plethysmogram, and respiration waveform. The factory default is “ECG”