Do not use batteries manufactured by other companies, which may cause damage to the device. If battery is
damaged, please replace with same type and specification battery marked by “CCC” or “CE” in time, or contact
the company directly.
Warning:
1. To avoid battery damage always remove battery(s) before shipping or storage.
2. It is recommended to use the battery specified by the manufacturer.
3. Life expectancy of a battery depends on how frequently and how long it is used. For a properly maintained and
stored lead-acid or lithium battery, its life expectancy is about 2 or 3 years respectively. For more aggressive use
mode, life expectancy can be less. We recommend replacing lead acid battery every 2 years and lithium battery
every 3 years.
Caution:
Keep the battery out of the reach of childrenNon smontare la batteria.
Do not disassemble battery
13.3 Battery Recycling
When a battery has visual signs of damage, or no longer holds a charge, it should be replaced. Remove the old battery
from the monitor and recycle it properly. To dispose of the batteries, follow local laws for proper disposal.
Chapter 14 Cleaning and Disinfection
14.1 Cleaning the Device and Accessories
Your device should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your place,
the device should be cleaned more frequently. Before cleaning the equipment, consult your hospital’s regulations for
cleaning the device.
Recommended cleaning agents are:
Sodium hypochlorite bleach (diluted)
Hydrogen peroxide (3%)
75% Ethanol
70% Isopropanol
To clean your device, follow these rules:
Switch off the monitor and disconnect the power cable before cleaning.
Kept the monitor from dust.
It is recommended to clean the outer shell and screen of the monitor to keep it clean. Only non-corrosive
cleanser such as clear water is permitted.
Wipe the surface of the monitor and transducers with an Ethanol impregnated wipe, and dry it with dry and