83
ENGLISH
3. RESP rate alarm limit setting range: High: 1rpm~150rpm; Low: 0rpm~149rpm.
4. Alarm tolerance: ±1rpm
17.3 TEMP
1. TEMP measuring range: 21.0
℃
~50.0
℃
2. TEMP measuring accuracy: ±0.2
℃
for the range from 25
℃
to 45
℃
, 0.4
℃
for other ranges.
3. TEMP responding time: ≤150s
4. TEMP minimum measuring time: ≥130s
5. Measuring site: body surface
6. Mode of operation: direct mode
7. Unit:
℃
and °F
8. TEMP alarm limit setting range: High: 0
℃
~60
℃
; Low: 0
℃
~59.9
℃
.
9. Tolerance: ±0.1
℃
17.4 NIBP
1. Measuring method: Oscillometric Technique
2. Pneumatic pressure measuring range: 0 mmHg~300mmHg
3. Accuracy of pressure measurement: ±3 mmHg
4. Typical measurement time: <30 seconds (adult cuff)
5. Measurement time on the average: < 90 seconds
6. Air release time while the measurement is canceled: <2 seconds (typical adult cuff)
7. Initial cuff inflation pressure
Adult: <150 mmHg; Infant: <120 mmHg; Neonate: <70 mmHg Tolerance: ±5 mmHg
8. Overpressure protection limit
Adult: 300 mmHg; Infant: 240mmHg; Neonate: 150 mmHg
9. NIBP measurement range:
Pressure
(
unit)
Adult
Infant
Neonate
SYS
mmHg
40~275
40~200
40~135
kPa
5.3
~
36.7
5.3
~
26.7
5.3
~
18.0
MAP
mmHg
20~230
20~165
20~110
kPa
2.7
~
30
2.7
~
22.0
2.7
~
14.6
DIA
mmHg
10~210
10~150
10~95
kPa
1.3
~
28
1.3
~
20.0
1.3
~
12.7
10. NIBP accuracy:
Maximum mean difference: ±5 mmHg
Maximum Standard deviation: 8 mmHg
11. Measurement mode: Manual, Auto, STAT
12. NIBP Alarm setting range:
see Section Alarms
17.5 SpO
2
1. Transducer: dual-wavelength LED
Wavelength: Red light: 660 nm, Infrared light: 905 nm