![Gima 35134 Operator'S Manual Download Page 237](http://html1.mh-extra.com/html/gima/35134/35134_operators-manual_2228002237.webp)
49
ENGLISH
6.4.1 High and Low Alarm Setting Range
Parameter
Setting range
High limit
Low limit
HR
(
bpm
)
(Low limit+1)~350
0~(High limit-1)
S-T
(
mV
)
(Low limit+0.01)~2.50
-2.5~(High limit-0.01)
SpO
2
(
%
)
(Low limit+1)~100
0~(High limit-1)
PR
(
bpm
)
(Low limit+1)~300
0~(High limit-1)
RR
(
rpm
)
(Low limit+1)~150
0~(High limit-1)
TEMP1
(℃)
(Low limit+0.1)~60.0
0~(High limit-0.1)
TEMP2
(℃)
(Low limit+0.1)~60.0
0~(High limit-0.1)
TD
(℃)
0.0~5.0
NIBP (Unit)
mmHg
Adult
Pediatric
Neonate
SYS
High limit
(
Low limit+1
)~
280
(
Low limit+1
)~
200
(
Low limit+1
)~
135
Low limit
29
~(
High limit-1
)
29
~(
High limit-1
)
29
~(
High limit-1
)
MAP
High limit
(
Low limit+1
)~
242
(
Low limit+1
)~
165
(
Low limit+1
)~
110
Low limit
20
~(
High limit-1
)
20
~(
High limit-1
)
20
~(
High limit-1
)
DIA
High limit
(
Low limit+1
)~
232
(
Low limit+1
)~
150
(
Low limit+1
)~
100
Low limit
10
~(
High limit-1
)
10
~(
High limit-1
)
10
~(
High limit-1
)
NIBP (Unit)
mmHg
Adult
Pediatric
Neonate
SYS
High limit
(
Low limit+0.1
)~
37.3
(
Low limit+0.1
)~
26.7
(
Low limit+0.1
)~
18.0
Low limit
3.9
~(
High limit-0.1
)
3.9
~(
High limit-1
)
3.9
~(
High limit-0.1
)
MAP
High limit
(
Low limit+0.1
)~
32.3
(
Low limit+0.1
)~
22.0
(
Low limit+0.1
)~
14.7
Low limit
2.7
~(
High limit-0.1
)
2.7
~(
High limit-0.1
)
2.7
~(
High limit-0.1
)