![Gima 35134 Operator'S Manual Download Page 220](http://html1.mh-extra.com/html/gima/35134/35134_operators-manual_2228002220.webp)
32
ENGLISH
4.5 Screen Display for Arrhythmia Event List
Figure 4.13 Arrhythmia Event List Screen
4.5.1 Operation Instructions
The structure is similar to the recall screen. We will cover each function key below.
Start
: This button is used to start and end the system Arrhythmia detection. The default is OFF. When the Arrhythmia is
not ON, the “Learn” key is disabled. Press this key and the monitor enters learning stage before detecting arrhythmia.
The “Start” changes to “End”. Press it again to end the learning. When “Learn” key change to yellow from gray, it indicates
the learning has finished. When the ARR detection begins after learning, the monitor will automatically detect the ARR
event. If an ARR event is detected, the ECG waveform with corresponding ARR event will be displayed in the 3
rd
trace of
waveform, shown in Figure 4.13.
When the monitor get Reset or the patient ID has been changed, the ARR detection needs to be re-learned.
Learn
:
Because the Arrhythmia detection is based on the pattern of normal ECG waveform, which is built up during
learning stage. When the patient changes, the ARR detection could be incorrect, so re-learning is needed. To get better
result of the learning function, it is recommended to start learning while a segment of good quality ECG waveform
appears during the ECG monitoring.
View
: Press this key, and the selected record in the Arrhythmia event list, which is to be reviewed, becomes the green.
Rotate the “Navigation Knob” to choose the record and display the corresponding waveform in the 3
rd
trace of waveform.
Press the knob again to exit.
Exit
: press this key to exit the Arrhythmia detection screen and return to the System Menu screen.