44
ENGLISH
Exit:
return to the System Setup screen.
5.6 SpO
2
Related Settings
On “System Menu” screen, select “Setup”
→
“SpO
2
” to enter into SpO
2
related settings.
Figure 5.6 SpO
2
related settings
SpO
2
Hi/Lo
: High and Low alarm limits for SpO
2
Pulse Hi/Lo
: High and Low alarm limits for pulse rate.
Exit:
return to the System Setup screen.
5.7 Respiration Related Settings
On “System Menu” screen, select “Setup”
→
“RESP” to enter into Respiration
related settings.
Figure 5.7 Respiration related settings
Gain
:
Respiration amplification/gain, 4 options, x1/2, x1, x2, and x4. The default is x2
Speed
:
Respiration waveform sweeping speed, 2 options 6.25mm/s and 12.5 mm/s. The default is 12.5 mm/s
Apnea
:
timeout setting for apnea alarm. Options: “OFF” and any number from 5 to 120. The step is 1 and its unit is
second. When the breathing action is not detected after the time longer than this set period, the trace for respiration
waveform will display warning “Apnea xxx second”. The default is “OFF”.