Chapter 8: System Hardware Overview
5495975-1EN Rev.9
8-60
© 2013-2017 General Electric Company. All rights reserved.
3. Table Top Positioning
Foot Pedals
This pedal allows you to move the table in all directions: longitudinally and
transversely. This is known as a floating tabletop.
Press the pedal twice and hold down to move the table top.
4. Down Pedal
The Down pedal lowers the table down.
Press the pedal twice and hold down to lower the table.
5. Up Pedal
The Up pedal raises the table up.
Press the pedal twice and hold down to raise the table.
6. Table Lock Control
Buttons
There are two lock buttons located on the right sides of the table, near the
head and foot of the table. They are used to prevent the table up-down
and tabletop movement. The lock is engaged when the button is lit.
7. Emergency Stop
Buttons
There are two Emergency Stop buttons located on the left sides of the
table. These buttons are used to power off the table in an emergency.
Refer to
Chapter 4: General Information-Emergency Stop (p. 4-9)
for more
information.
8. Detector Tray
Override
The Detector Tray Override is used for receptor placement if auto-tracking
fails.
9. Optical pinch sensor
area
The sensors detect a hand or other body part below the surface of the
table and lock the table movement to prevent injury. The system will sound
an audible tone when the finger pinch lock is activated.
Refer to
Digital Cassette Imaging in Extended Table Detector Tray(p. 8-56)
for more information.
Item
Description