Chapter 13: Advanced Applications
5495975-1EN Rev.9
13-12
© 2013-2017 General Electric Company. All rights reserved.
Follow this process to export RRA data:
1. On the Worklist screen, click [UTILITIES].
2. Select RRA Reporting Tool [START] button.
3. Set the
From
date and time
4. Set the
To
date and time
5. Insert a blank CD-R into the disk tray or connect a USB (minimum 64 MB) device to a USB port on the
acquisition workstation.
IMPORTANT!
The system cannot export RRA data to CD-RW or DVD.
6. Click [EXPORT TO CD] or [EXPORT TO USB] as appropriate.
A message appears: “Number of images in report: __
Press Continue to write on CD or USB.
Press Cancel to cancel export.”
7. Click [CONTINUE].
8. Wait for the export to complete.
9. Remove the CD or USB device.
10. Click [CLOSE].
Not Enough Space to Save RRA Images
If the disk or USB device does not have enough storage space to save all the RRA images, the following
message will appear. “RRA data exceeds media capacity.
Select Continue to export without JPEGs.
Select Cancel to cancel export.”
Clicking [CONTINUE] will export the Report and Data spreadsheet only. Images will not be saved.
There are two options to retain images:
Obtain a USB device with larger storage capacity and repeat the steps to export the data.
Reduce the export interval and complete multiple exports to additional disks or USB devices.
Follow this process to save the JPEG data to additional disks or USB devices.
1. Click [CANCEL] to stop the export.
2. Obtain additional disks or USB devices.
3. Change the
To
date and time so that the report interval is shorter.
4. Click [EXPORT TO CD] or [EXPORT TO USB] as appropriate.
5. Click [CONTINUE] on the message to export the data.
If you still do not have enough space, click [CANCEL] to stop the export and change the
To
date to
create an even shorter interval.