Chapter 8: System Hardware Overview
5495975-1EN Rev.9
8-22
© 2013-2017 General Electric Company. All rights reserved.
Axes Indicators
The Optima XR646 shows axes indicators on the carriage (
). These axes markers are color
coded to match the release controls on the OTS user interface. The colors for the 3 axes are as follows:
Blue - Vertical axes
Green - Longitudinal axes
Pink - Lateral axes
Figure 8-10
Axes indicators and corresponding OTS release controls
Vertical Detent
There is a calibrated vertical detent for the OTS column that sets the locks when the X-Ray tube is at
100cm (40 inches) above the receptor or film. This detent position is selectable at system installation.
Longitudinal (green)
Lateral (pink)
Vertical (blue)