Chapter 8: System Hardware Overview
5495975-1EN Rev.9
8-14
© 2013-2017 General Electric Company. All rights reserved.
3. Indicator
Manual
Collimation
LED
Indicates when the system has switched to manual collimation
mode (orange steady light). In this mode, the collimator field of
view (FOV) is not limited to the receptor area. If the key switch is in
OVERRIDE position, the manual collimation LED flashes (yellow
blinking light).
4. Indicator
SID Detent
LED
Indicates configurable SID detents.
SID detents are (Wallstand is 0 degree): 102cm, 107cm, 112cm,
117cm, 183cm, and 2 custom SID values. The default SID is 183cm
(72inch).
SID detents are(Wallstand is 90 degree):
102cm, 107cm,112cm, 117cm, 127cm, and 2 custom SID values.
The default SID is 102cm (40inch).
1. SID detents will be set by Field Engineer accordiong to
customer’s request.
2. When reach to a SID detent the OTS would be locked along the
SID direction and this green light is on.
5. Indicator
Lateral
Detent LED
Indicates the lateral detent position.
When reach the lateral detent, the OTS would be locked along the
lateral direction and this green light is on.
6. Indicator
Longitudina
l Detent LED
Indicates the longitudinal detent position.
When reach the longitudinal detent, the OTS would be locked
along the longitudinal direction and this green light is on.
Table 8-6
OTS Function
Item Number and
Type
Description
Description