Chapter 10: Image Acquisition
5495975-1EN Rev.9
10-6
© 2013-2017 General Electric Company. All rights reserved.
mAs
Adjusts the mAs.
The mAs selection is in Renard steps. The available
selections are: 0.1, 0.12, 0.16, 0.20, 0.25, 0.32, 0.40, 0.50,
0.64, 0.80, 1.0, 1.25, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.4, 8.0, 10.0,
12.5, 16.0, 20.0, 25.0, 32.0, 40.0, 50.0, 64.0, 80.0, 100.0,
125.0, 128.0, 160.0, 200.0, 250.0, 256.0, 320.0, 400.0,
500.0, 512.0.
Note:
Not all mAs selections are available at all kV set-
tings.
If in AEC mode, shows the calculated mAs for the current
kV and mA after exposure.
mSec
Shows the exposure time for the technique with the
current kV, mA, and mAs settings after exposure is
completed.
If in AEC mode, the AEC back-up time is displayed below
the Sec field.
The AEC default backup time is two (2) seconds.
Focal Spot:
Selects a large or small focal spot.
CU Filtration:
Selects the amount of copper filtering. The selectable
range is 0.0mm, 0.1mm, or 0.2mm.
The recommended amount of filtering is shown below the
selection area.
Patient Side:
If conducting an exam on paired anatomy (for example,
extremities), selects the side of the patient being x-rayed.
Options are:
Both
Left
Right
If conducting an exam on unpaired anatomy, the control
is disabled and displays “Unpaired”, as shown here.
Table 10-1
Image Acquisition functions
Function
Description