Detector System List
Figure 15-15
Detector System List
The Detector System List contains the names of the Systems the Digital Detector has been registered
with. The Digital Detector can be registered with up to 20 systems.This list is contained in the memory of
the Digital Detector and goes with the Digital Detector from system to system.
Detector Sleep Time
Figure 15-16
Detector Sleep Time
The digital detector can be set to transition to sleep mode at a designated time to save the battery power
of the detector. If the detector sleep time is set to 15 minutes, it will transition into sleep mode 15 minutes
after the last exposure on the system. To take the detector our of sleep mode, press the black button on
the handle of the detector for one to two seconds. The detector will be ready for exposures within 10
seconds.
The available increments are 15, 30, 45mins and Never.