Chapter 10: Image Acquisition
5495975-1EN Rev.9
10-20
© 2013-2017 General Electric Company. All rights reserved.
Note:
If you are performing with more than one detector, please avoid the detectors overlapped
position when taking exposure. Because, with one detector placed in the table housing, when
another detector is positioned on the tabletop to take exam, the Exposure Inhibition may happen
when the two detectors overlapped. To dissolve the Exposure Inhibition, please move away any
one of the two detectors to avoid the overlap position.
7. Confirm or adjust the
Patient Side,
if applicable.
8. Collimate and shield as appropriate for the exam.
9. Have the patient suspend respiration, if required.
10. Make exposure using the hand switch.
The image appears on Image Viewer screen.
Conduct a Cassette Exam
The Cassette Exam function takes an x-ray without digital image acquisition and storage or electronic
information sending and retrieval. This is for use with a traditional Film Screen or CR cassette.
When the cassette receptor is selected, a message appears near the bottom of the screen: “CAUTION:
Non-digital Exposure. Ensure that film is in place.“