Chapter 2: Safety and Regulatory
5495975-1EN Rev.9
2-12
© 2013-2017 General Electric Company. All rights reserved.
2. Electromagnetic Immunity
Table 2-4
Guidance and manufacturer’s declaration – electromagnetic immunity
NOTE:
UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Optima XR646 is intended for use in the electromagnetic environment specified below. The customer or the user of the Optima
XR646 should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level Electromagnetic Environment Guidance
Electrostatic discharge
(ESD), IEC 61000-4-2
+/
−
6 kV contact.
+/
−
8 kV air.
+/
−
6 kV contact.
+/
−
8 kV air.
Floors should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/
burst, IEC 61000-4-4
+/
−
2 kV for power supply
lines.
+/
−
1 kV for input/output
lines.
+/
−
2 kV for power
supply lines.
+/
−
1 kV for input/
output lines.
Mains power quality should be that
of a typical commercial or hospital
environment
Surge, IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines, IEC 61000-4-11
< 5% UT (> 95% dip in UT)
for 0.5 cycle.
40% UT , (60% dip in UT)
for 5 cycles.
70% UT , (30% dip in UT)
for 25 cycles.
< 5% UT, (> 95% dip in
UT) for 5 cycles
< 5% UT (> 95%
dip in UT) for 0.5
cycle.
40% UT , (60% dip
in UT)for 5 cycles.
70% UT , (30% dip
in UT)for 25
cycles.
< 5% UT, (> 95%
dip in UT) for
5 cycles
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the system requires continued
operation during power mains interruptions, it
is recommended that the system be powered
from an un-interruptible power supply or a
battery.
Power frequency (50/60
Hz) magnetic field,
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.