Falcon 4M Camera Manual
Teledyne DALSA
03-032-20044-03
83
Command
Syntax Parameters
Description
set vid eo m od e
svm
m
Sets the camera‘s vid eo mod e.
0
: Vid eo m od e
1: Test p attern checkerboard
2
: Test p attern alternating line 1
3
: Test p attern alternating line 2
4
: Test p attern horizontal ram p
5
: Test p attern vertical ram p
6: Test p attern d iagonal ram p
7: Test p attern FPN
8: Test p attern PRN U
9: Test p attern fixed 1023
10: Test p attern PRN U m ap
11: Test p attern fixed 0
12: Test p attern FPN m ap
test p attern valu e
tpv
i
Sets the base valu e (63/ 127/ 255) for
svm 10. A PRN U m ap resu lts w hen this
base valu e is m u ltip lied by ind ivid u al
PRN U coefficients.
verify tem p eratu re
vt
Disp lay the internal tem p eratu re of the
cam era.
verify voltage
vv
Disp lay som e internal voltages
su p p lied to the cam era.
w rite fp n
coefficients
wfc
Write cu rrent FPN coefficients to non -
volatile m em ory. The set w ithin non-
volatile m em ory w ill have been
p reviou sly selected u sing the
csn
com m and .
w rite p rn u
coefficients
wpc
Write cu rrent PRN U coefficients to non -
volatile m em ory. The set w ithin non-
volatile m em ory w ill have been
p reviou sly selected u sing the
csn
com m and .
Parameters:
t
= tap id
i
= integer value
f
= real number
s
= string
m
= member of a set