Test Functions
340
MAX-800 Series
OH - OTN
Row 2
RES
, columns 1-2, ODU OH: The two Reserved (RES) bytes are
configurable from
00
(default for each byte) to
FF
.
PM & TCM
, column 3, ODU OH: The Path Monitoring and Tandem
Connection Monitoring byte is configurable from
00
(default) to
FF
.
TCM ACT
, column 4, ODU OH: The Tandem Connection Monitoring
Activation is configurable from
00
(default) to
FF
.
TCM6
/
TCM5
/
TCM4
, column 5-13, ODU OH: The Tandem Connection
Monitoring overhead contains the following bytes.
The first TCMi byte contains the TTI multiframe byte and is only
configurable from
The second TCMi byte contains the BIP-8 byte and is automatically
generated for each frame. This byte is not configurable.
The third TCMi byte contains the following sub-fields. This byte is
configurable from
00
to
FF
. The default value is
00
when TCMi is
disabled, and 01 when enabled.
FTFL
, column 14, ODU OH: The Fault Type Fault Location multiframe
byte is only configurable from
RES
, column 15, OPU OH: The Reserved (RES) byte is configurable
from
00
(default) to
FF
.
JC
, column 16, OPU OH:
Bits 1-6, Reserved (RES), are configurable from binary
000000
(default) to
111111
.
Bits 7-8, Justification Control (JC), are configurable from binary
00
(default) to
11
. Not available with ODU mux. Changing the JC value
will corrupt the payload.
Bit 1-4
Bit 5
Bit 6-8
BEI/BIAE
BDI
STAT
Summary of Contents for MAX-800 Series
Page 1: ...www EXFO com User Guide MAX 800 Series MaxTester Application ETHERNET AND TRANSPORT TESTER...
Page 16: ......
Page 18: ......
Page 26: ......
Page 43: ...Test Setup Test Applications MaxTester 35 SONET SDH BERT For SDH BERT...
Page 48: ...Test Setup Test Applications 40 MAX 800 Series SONET SDH DSn PDH BERT For SDH...
Page 62: ......
Page 208: ......
Page 414: ......
Page 426: ......
Page 473: ...Glossary MaxTester 465 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values...
Page 476: ......
Page 494: ...0 5 1 5 48 5 0 176 W d y K y K D...