Test Results
268
MAX-800 Series
Performance Monitoring
For
M.2100 ISM
: Gives the count of the seconds within which
anomalies (frame bit errors, CRC block errors, etc.) are
Y or at least
one defect occurred. Y depends on the type of DSn/PDH signal as
described in the following table.
BBE
(Background Block Error) (
G.826
,
G.828
,
G.829
, and
M.2101
):
Gives the count of Errored Block not occurring as part of a SES.
Signal
SES Threshold
DS1 (SF)
8 frame bit errors (Near-End)
DS1 (ESF)
320 CRC-6 block errors (Near-End) 320 CRC-6 block errors (Far-End, if FDL
enabled)
E1 (Framed without CRC-4) 28 frame bit errors (Near-End)
E1 (Framed with CRC-4)
805 CRC-4 block errors (Near-End) 805 E-bit errors (Far-End)
DS3 (M13)
2444 P-bit errors (Near-End) or 5 F-bit errors (Near-End)
DS3 (C-bit Parity)
2444 P-bit errors (Near-End) or 5 F-bit errors (Near-End) 2444 FEBE errors
(Far-End)
E2 (Framed)
41 frame bit errors (Near-End)
E3 (Framed)
52 frame bit errors (Near-End)
E4 (Framed)
69 frame bit errors (Near-End)
Summary of Contents for MAX-800 Series
Page 1: ...www EXFO com User Guide MAX 800 Series MaxTester Application ETHERNET AND TRANSPORT TESTER...
Page 16: ......
Page 18: ......
Page 26: ......
Page 43: ...Test Setup Test Applications MaxTester 35 SONET SDH BERT For SDH BERT...
Page 48: ...Test Setup Test Applications 40 MAX 800 Series SONET SDH DSn PDH BERT For SDH...
Page 62: ......
Page 208: ......
Page 414: ......
Page 426: ......
Page 473: ...Glossary MaxTester 465 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values...
Page 476: ......
Page 494: ...0 5 1 5 48 5 0 176 W d y K y K D...